Skip to main content

Table 1 AMH intra- and inter-assay precision on Access2 and DxI800 analysers

From: Assessment of the Access AMH assay as an automated, high-performance replacement for the AMH Generation II manual ELISA

  

Intra-assay precision

Inter-assay precision

  

Mean, pmol/L

SD, pmol/L

CV, %

Mean, pmol/L

SD, pmol/L

CV, %

Access2

Control 1

6.27

0.09

1.51

6.23

0.24

3.85

 

Control 2

31.78

0.45

1.41

31.69

0.96

3.04

 

Control 3

97.85

1.64

1.68

97.19

3.35

3.45

DxI800

Control 1

6.40

0.21

3.30

6.31

0.28

4.36

 

Control 2

33.36

0.68

2.04

32.49

1.87

5.76

 

Control 3

103.18

2.98

2.89

98.54

4.52

4.59