Skip to main content

Table 3 Comparison of different dilution methods using the Access AMH Assay

From: Assessment of the Access AMH assay as an automated, high-performance replacement for the AMH Generation II manual ELISA

  

Dilution in patient sera

Dilution in sample diluent A

 

Dilution

Mean, pmol/L

CV, %

Deviation, %

Mean, pmol/L

CV, %

Deviation, %

Sample 1

Neat

76.90

0.50

0.00

76.90

0.50

0.00

 

1:5

74.82

1.09

−2.71

68.90

1.73

−10.41

 

1:10

74.63

1.59

−2.96

66.28

2.74

−13.81

 

1:16

77.81

1.11

1.18

68.04

1.36

−11.52

Sample 2

Neat

170.64

1.83

0.00

170.64

1.83

0.00

 

1:5

177.69

1.10

4.13

164.44

1.24

−3.63

 

1:10

176.71

0.64

3.56

154.80

0.46

−9.28

 

1:16

179.53

3.55

5.21

153.65

3.07

−9.96